                  當(dāng)前位置:首頁 > 新聞中心 > U23分析儀故障維修(中英文對照)

                  U23分析儀故障維修(中英文對照)

                  fault message(故障信息) possible causes(可能的原因) remedy(補(bǔ)救辦法)

                  flow too low during measuring (在測量流量太低)

                  1:sample gas blocked or leaky(樣品氣堵塞或泄漏)

                  A:clean or replace the blocked parts(hose filter etc) if the problem is still present(清潔或更換堵塞的配件(軟管過濾器等)如果問題仍然存在)

                  2:pump not running(泵沒有運(yùn)行)

                  A:switch the pump on(開關(guān)泵)

                  B:pump capacity too low(泵的容量太低)

                  C:increase the pump capacity as described in section(提高泵的容量中所描述的部分)

                  D:pump faulty(泵故障)

                  no data for temperature compensation(沒有數(shù)據(jù)進(jìn)行溫度補(bǔ)償)

                  1:temperature compensation not completed successfully(溫度補(bǔ)償沒有成功完成)

                  2:new components have been loaded(新的組件被加載)

                  3:EEPROM has been initialized(EE PROM已初始化)flow too low during autocal(在autocal流量太低)

                  A:cal gas path blocked or leaky(校準(zhǔn)氣路堵塞或泄漏)

                  B: clean or replace the blocked parts (hose filter etc) if the problem is still present(清潔或更換堵塞的配件(軟管過濾器等)如果問題仍然存在)

                  conc of O2 too low during autocal (濃度的氧氣在autocal太低)

                  1:O2 sensor faulty or old(氧傳感器的故障或老化)

                  A:replace the O2 sensor as described in section(更換氧傳感器中描述的部分)meas value display( 測量值顯示)

                  2:O2 sensor zero not calibrated(氧傳感器的零點(diǎn)未校準(zhǔn))

                  calibrate the zero of the o2 sensor as described in section(校準(zhǔn)零的氧氣傳感器中描述的部分)fault at analog output(故障模擬輸出)

                  A:component could not be initialized when switching on(組件無法初始化時開關(guān))

                  B:limits were violated when calibrating the analog section(限制受到侵犯時,校準(zhǔn)模擬部分)general fault of all ir channels mv-c1/2/3 1:chopper faulty(斬波器故障)beyond tolerance(一般故障的紅外通道m(xù)v-c1 / 2 / 3超出公差)

                  IR source faulty(紅外源錯誤)fault of addresses for ir channels(故障紅外通道地址)

                  1:plug-in jumper on detector for detection of components not ok(插件在探測器檢測元件不好)

                  2:the cable of the detector has no contact(電纜探測器沒有接觸)

                  A:check that the plug is correctly connected to the detector (the plug must latch in twice)(檢查插頭是否正確連接到探測器(插頭鎖在兩次))

                  B:cable of detector faulty(電纜故障探測器)meas value channel 1 beyond tolerance(測量值超出公差1頻道)

                  analyzer section of first component is faulty(分析儀部分*組件是錯誤的)meas value display:*****(測量值顯示:*****)

                  meas value o2 beyond tolerance(測量氧氣超出公差值)

                  1:replace the O2 sensor as described in section(更換氧傳感器中描述的部分)meas value display:****(測量值顯示:* * * *)

                  2:motherboard faulty(主板故障)mains voltage beyond tolerance(電源電壓超出公差)

                  1:mains voltage varies(電源電壓變化)

                  A:correct such that the mains voltage is within the tolerance values permissible for the analyzer(正確的電源電壓范圍內(nèi)允許的公差值分析儀)

                  power spply unit faulty(供電單元故障)temp of analyzer beyond tolerance(溫度超出公差分析)

                  1:ambient temperature too high or too low(環(huán)境溫度過高或過低)

                  A:provide sufficient ventilation or air conditioning(提供充分的通風(fēng)或空調(diào))

                  2:temperature sensor of a detector faulty(溫度傳感器探測器故障)pressure of amb air beyond tolerance(壓力的空氣超出公差)

                  A:pressure sensor faulty(壓力傳感器故障)autocal drift beyond tolerance(autocal漂移超出公差)
                  西門子U23氣體分析儀維修(7MB2335、7MB2337、7MB2338系列)
                   歡迎詢價:
                  詢價號:2079705868

                  注:發(fā)傳真時請注明張工收,我給您低的價格

                  上海湘乾儀器儀表有限公司(m.j9001j.com)關(guān)鍵字:西門子氣體分析儀,ULTRAMAT23氣體分析儀,西門子LDS6激光分析儀

                  上海湘乾儀器儀表有限公司 版權(quán)所有 總訪問量:254191

                  郵箱:13391236317@163.com GoogleSitemap 管理登陸 ICP備案號:

                  儀表網(wǎng)

                  推薦收藏該企業(yè)網(wǎng)站
                  精品久久久一区二区_秋霞久久精品_日日干天天摸_日本午夜福利影院_来一水AV@lysav
                  • <small id="lqd1h"></small>
                    1. <noscript id="lqd1h"><dl id="lqd1h"></dl></noscript>
                        <small id="lqd1h"><tbody id="lqd1h"></tbody></small>
                      1. 
                        
                      2. <i id="lqd1h"><ins id="lqd1h"></ins></i>
                                滨海县| 兴海县| 鄂州市| 盐池县| 绥宁县| 萨嘎县| 中宁县| 肇庆市| 昌图县| 黄梅县| 上饶县| 南投市| 茂名市| 游戏| 嘉禾县| 乃东县| 如东县| 论坛| 务川| 万山特区| 郯城县| 诏安县| 汉寿县| 嘉兴市| 三江| 涪陵区| 贵溪市| 海淀区| 广水市| 山西省| 沙田区| 北宁市| 商河县| 洪洞县| 行唐县| 孟村| 彭山县| 桦川县| 大田县| 临沭县| 仙居县|